Paper tube compression tester operating instructions

Paper tube compression tester operating instructions

Dongguan Haida Instrument 400-071-1808/0769-89280808

The system is calibrated after it is installed in the testing machine (only need to perform this operation before the first test).

Before the test begins, you need to set the test parameters. Otherwise, the system will use the last parameter or the default parameter by default.

Note: After calibration and after setting parameters, please pay attention to check and save the data ( will be saved automatically under some interfaces ) . The system will automatically use the saved parameters at the next startup.

Gain 1 is related to the sensitivity of the sensor and can be matched with hardware's dial switch;

Make gain 1 in place. The position of the dial switch is changed and the hardware zero should be recalibrated.

1 Select the motor type and select the correspondence between the direction of movement of the machine and the force applied to the sensor (must be operated by a professional).

2 Calibrate the speed.

Calibration method: The speed calibration sequence is from the principle of large to small, select the appropriate speed, adjust the parameters, so that the motor speed is within the operating error range. If the DC motor is 100mm/min or less, it must be calibrated point by point, and 100 or more can be calculated based on 500mm/min . If the stepper or servo motor is calibrated with a single speed, the calibration can be completed.

3 Calibration load

Calibration steps:

A ) Enter the sensor configuration to match the configuration information with the actual value

B ) Adjust the hardware zero to display 0 .

C ) Enter the gain 1 adjustment, the load is loaded to full scale ( 80% to 100% ) × 100 %, so that the displayed value is consistent with the actual value

   Note: Gain 1 is the base gain, other gains are trim gains and are based on gain 1 .

D ) Enter the gain 2 adjustment, the load is loaded to the full scale ( 80% to 100% ) × 50 %, so that the displayed value is consistent with the actual value

E ) Enter the gain 3 adjustment, the load is loaded to the full scale ( 80% to 100% ) × 20 %, so that the displayed value is consistent with the actual value

F ) Enter the gain 4 adjustment and the load is loaded to the full scale ( 80% to 100% ) × 10 % so that the displayed value is consistent with the actual value.