Digital display shaking incubator manual

Models of digital display shaking incubator: BS-1E BS-2F BS-4G
Oscillation incubator is a biochemical instrument with a heating and cooling bidirectional temperature regulation system, a temperature-controlled incubator and an oscillator. It is a plant, biology, microorganism, genetics, viral medicine, environmental protection, food, petroleum, chemical industry, etc. The scientific research, education and production departments make indispensable laboratory equipment for precision cultivation.
1. Main features of digital display oscillation incubator:
1. The heat insulation material of the box body is made of polyurethane foam foamed on site, which has strong anti-interference ability to external cold (heat) sources.
2. There is an air duct in the working chamber, and the temperature distribution is even.
3. The inner wall is made of stainless steel with good corrosion resistance.
4. The heating system can raise the temperature to 50 ℃ when the environment is -5 ℃.
5. The refrigeration system can be reduced to 5 ℃ when the ambient temperature is 32 ℃.
6. The universal spring test bottle rack is particularly suitable for the cultivation and preparation of biological samples for various comparison tests.
7. Stepless speed regulation, safe operation.
8. Precise temperature control, LED digital display.
Second, the digital display oscillation incubator technical instructions:

Index \ Model
BS-1E
BS-2F
BS-4G
Volume
140L
280L
420L
Temperature control range
5-50 ℃
5-50 ℃
5-50 ℃
Temperature control accuracy
± 1 ℃
± 1 ℃
± 1 ℃
Oscillation speed
Start-300r / min
Start-300r / min
Start-300r / min
amplitude
20mm (swing)
20mm (swing)
20mm (swing)
Number of oscillation groups
1
2
4
Filling volume
Test tube: ¢ 16 × 200
100ml × 15 200ml × 9
Test tube: ¢ 16 × 200
100ml × 15 × 2 200ml × 9 × 2
Test tube: ¢ 16 × 200
100ml × 15 × 4 200ml × 9 × 4
heating power
300W
400W
600W
Compressor power
130W
180W
240W
power supply
AC 220V50Hz
AC 220V50Hz
AC 220V50Hz
Dimensions
610 × 610 × 1150 (mm)
1250 × 610 × 1460 (mm)
1440 × 710 × 1770 (mm)
3. Instructions for use of digital display incubator:
1. The instrument should be placed on a flat ground, the environment should be clean and tidy, dry and ventilated.
2. Before the instrument is used, each control switch should be in a non-working state, and the speed control knob should be placed in the minimum position.
3. When cultivating the test bottles, in order to make the instrument have a good balance performance when oscillating and avoid generating large vibrations, all the test bottles should be filled up when bottling, and the culture solution of each bottle should be approximately equal. If the culture medium is insufficient, the test bottles can be placed symmetrically.
4. Turn on the external power supply, set the power switch to the "on" position, and the indicator light is on.
5. Select the cultivation temperature: please refer to the instruction manual.
6. Turn on the "oscillation switch, the indicator light is on, and slowly adjust the speed control knob to the required speed.
The medium-speed use within the speed adjustment range can extend the service life of the instrument.
7. Before each stop, each control switch should be in a non-working state, cut off the power supply.
8. Matters needing attention:
A: The incubator of two groups of oscillators, the oscillators are controlled by the respective "oscillation switch" and "speed control knob".
B: It is not suitable for the cultivation of articles containing volatile chemical solvents, low-concentration explosive gas and low ignition point gas, and toxic substances.
1. Medium-speed use within the speed range can extend the service life of the instrument.
2. The instrument should be placed on a firmer workbench, and the environment should be clean and tidy with good ventilation.
3. When placing items on the workbench, appropriate intervals should be maintained between the test bottles to facilitate convection circulation of cold and hot air.
4. A large glass door is used to better observe the changes in the oscillating container. If transparency is not required, black paper can be attached with transparent tape.
5. Be optimistic about the oscillator and adjust it to the speed you need. Type A has digital display speed.
6. Observe the working cavity and turn on the lighting switch.

4. Maintenance of digital display oscillation incubator:
1. Properly use and pay attention to the maintenance of the instrument to make it in good working condition, which can extend the service life of the instrument.
2. After the refrigeration system stops working, wipe the working chamber and glass observation window with a soft cloth.
During the continuous working period of the instrument, periodic inspection should be done every three months; check whether there are water drops, dirt, etc. falling on the motor and exposed refrigeration components; clean the dust and dirt on the compressor, condenser; Control elements and fastening screws.
3. The bearings of the drive part of the oscillator of the digital display incubator have been filled with a proper amount of grease (No. 1 calcium-sodium base) before leaving the factory. During the continuous operation of the instrument, the grease should be filled every six months and filled The volume accounts for about 1/3 of the bearing space.
4. After long-term use of the instrument, natural wear is normal. After the instrument is used for one year, if the motor, compressor have abnormal noise, cooling, heating system abnormal transmission mechanism bearing wear, electronic control component failure ... and other failures, the company will continue to provide quality services to assist deal with.
From: http: //