Common problems with ultrasonic transducers

Common Problems of Ultrasonic Transducer 1. The ultrasonic vibrator is damp. You can use a megohm meter to check the plug connected to the transducer. You can judge the basic situation by checking the insulation resistance value. Generally, the insulation resistance is more than 5 megohms. If this insulation resistance value is not reached, generally the transducer is damp, you can put the whole transducer (excluding the plastic shell) into the oven and set it to dry at about 100 ℃ for 3 hours or use a hair dryer to remove the damp to the resistance value So far.
2. The transducer vibrator ignites and the ceramic material is cracked. It can be checked with the naked eye and a megohmmeter. It is generally used as an emergency treatment measure. It can disconnect individual damaged vibrators and will not affect the normal use of other vibrators.
3. The vibrator is degummed. Our transducer adopts the double guarantee process of cementing and screw fastening, which will not happen under normal circumstances.
4. Perforation of the vibration surface of stainless steel. Generally, if the transducer is used at full load for 10 years, the perforation of the vibration surface may occur.

Ultrasonic Cell Crusher / Ultrasonic Cell Cracker / Ultrasonic Cell Crusher / Ultrasonic Cell Crusher

http: //

Makeup for Eyes, those products mainly include eye liner, Mascara, Eyeshadow, and other related product.

Its mainly for protect the eyes. Its looks more beautiful and shining after the product using, also can keep the long lasting and waterproof.

26

Eyes

Color Makeup for Eyes, Color Eyeshadow, Gel Eyeliner, Liquid Eyeliner, Mascara, Eyebrow Gel

Zhejiang Zhongyimei Industry Co., Ltd , https://www.queeeny.com