Reovirus (RV) ELISA kit instruction manual

Reovirus (RV) ELISA kit instruction manual
This reagent is for research use only
Experimental principle:
This kit uses double antibody sandwich enzyme-linked immunoassay (ELISA) to determine the reovirus (RV) in the specimen. The microplate is coated with purified reovirus (RV) antibody to make a solid-phase antibody, which can be combined with reovirus (RV) in the sample. After washing to remove unbound antigen and other components, it is combined with The HRP-labeled reovirus (RV) antibody binds to form an antigen-antibody-enzyme-labeled antigen complex. After thorough washing, the substrate TMB is added for color development. TMB is converted into blue under the catalysis of HRP enzyme, and into the final yellow under the action of acid. The absorbance (OD value) was measured with a microplate reader at a wavelength of 450 nm and compared with the CUTOFF value to determine the presence or absence of reovirus (RV) in the specimen.
Kit composition:
Kit composition
48 hole configuration
96-well configuration
save
Instructions
1 serving
1 serving
Sealing film
2 pieces (48)
2 pieces (96)
sealed bag
1
1
Enzyme coated plate
1 × 48
1 × 96
Store at 2-8 ℃
Negative control
0.5ml × 1 bottle
0.5ml × 1 bottle
Store at 2-8 ℃
Positive control
0.5ml × 1 bottle
0.5ml × 1 bottle
Store at 2-8 ℃
Enzyme reagent
3 ml × 1 bottle
6 ml × 1 bottle
Store at 2-8 ℃
Sample diluent
3 ml × 1 bottle
6 ml × 1 bottle
Store at 2-8 ℃
Developer A liquid
3 ml × 1 bottle
6 ml × 1 bottle
Store at 2-8 ℃
Developer B liquid
3 ml × 1 bottle
6 ml × 1 bottle
Store at 2-8 ℃
Stop solution
3ml × 1 bottle
6ml × 1 bottle
Store at 2-8 ℃
Concentrated washing liquid
(20ml × 20 times) × 1 bottle
(20ml × 30 times) × 1 bottle
Store at 2-8 ℃
Sample processing and requirements:
1. Serum: room temperature blood coagulates naturally for 10-20 minutes, centrifuged for about 20 minutes (2000-3000 rpm). Collect the supernatant carefully and centrifuge again if a precipitate appears during storage.
2. Plasma: EDTA or sodium citrate should be selected as the anticoagulant according to the requirements of the specimen, mixed for 10-20 minutes, and centrifuged for about 20 minutes (2000-3000 rpm). Collect the supernatant carefully. If a precipitate forms during storage, it should be centrifuged again.
3. Urine: collected in a sterile tube and centrifuged for about 20 minutes (2000-3000 rpm). Collect the supernatant carefully. If a precipitate forms during storage, centrifuge again. Pleural and ascites, cerebrospinal fluid reference implementation.
4. Cell culture supernatant: When detecting secreted components, collect with a sterile tube. Centrifuge for about 20 minutes (2000-3000 rpm). Collect the supernatant carefully. When detecting the components inside the cells, dilute the cell suspension with PBS (PH7.2-7.4), and the cell concentration will reach about 1 million / ml. Through repeated freezing and thawing, the cells are destroyed and the intracellular components are released. Centrifuge for about 20 minutes (2000-3000 rpm). Collect the supernatant carefully. If a precipitate forms during storage, it should be centrifuged again.
5. Organize the specimen: after cutting the specimen, weigh it. Add a certain amount of PBS, PH7.4. Quickly freeze and save with liquid nitrogen for later use. After the specimen melts, it still maintains a temperature of 2-8 ° C. Add a certain amount of PBS (PH7.4) and homogenize the specimen with a manual or homogenizer. Centrifuge for about 20 minutes (2000-3000 rpm). Collect the supernatant carefully. After aliquoting, a portion is to be tested, and the rest is frozen for future use.
6. The specimen should be extracted as soon as possible after collection. The extraction should be carried out according to relevant literature. The experiment should be carried out as soon as possible after extraction. If the test cannot be performed immediately, the specimen can be stored at -20 ℃, but repeated freezing and thawing should be avoided.
7. The sample containing NaN3 cannot be detected because NaN3 inhibits the activity of horseradish peroxidase (HRP).
Steps:
1. Numbering: serially number the corresponding microwells of the sample, each plate should be set with 2 wells for negative control, 2 wells for positive control, and 1 well for blank control (the blank control well does not add sample and enzyme reagent, the rest of the steps are the same)
2. Add sample: add negative control and positive control 50μl to the negative and positive control wells respectively. Then add 40μl of sample diluent to the sample well, and then add 10μl of the sample to be tested. Add the sample to the bottom of the well of the microplate, try not to touch the wall of the well, gently shake to mix,
3. Incubation: seal the plate with a sealing film and incubate at 37 ° C for 30 minutes.
4. Mixing solution: Add 30 times (20 times of 48T) concentrated washing liquid to 600ml with distilled water and set aside
5. Washing: Carefully peel off the sealing film, discard the liquid, spin dry, fill each well with the washing liquid, let it stand for 30 seconds and then discard, repeat this 5 times, pat dry.
6. Add enzyme: add 50μl of enzyme label reagent to each well, except blank well.
7. Incubation: The operation is the same as 3.
8. Washing: The operation is the same as 5.
9. Color development: add 50μl of developer A to each well, then add 50μl of developer B, mix gently, and develop at 37 ℃ for 15 minutes in the dark
10. Termination: Add 50μl of stop solution to each well to stop the reaction (at this time, the blue color turns to yellow).
11. Determination: Measure the absorbance (OD value) of each well in sequence with the blank air conditioner at zero and 450 nm wavelength. The measurement should be carried out within 15 minutes after adding the stop solution.
Result judgment:
Test effectiveness: the average of positive control wells ≥1.00; the average of negative control wells ≤0.10
Calculation of critical value (CUT OFF): critical value = average value of negative control wells + 0.15
Negative judgment: samples with OD value <cut-off value (CUT OFF) are negative for reovirus (RV)
Positive judgment: the sample with OD value ≥ critical value (CUT OFF) is positive for reovirus (RV)

Precautions
1. The operation is strictly in accordance with the instructions. The components of different batches of this reagent must not be mixed.
2. The kit should be equilibrated at room temperature for 15-30 minutes before being taken out of the refrigerated environment. If the enzyme label coated plate is unopened, the strip should be stored in a sealed bag.
3. Crystals may be precipitated in the concentrated washing liquid, which can be heated and dissolved in a water bath during dilution, and the results will not be affected during washing.
4. The sealing film is limited to one-time use to avoid cross-contamination.
5. Please keep the substrate away from light.
6. The test results must be determined based on the reading of the microplate reader. When using dual wavelength detection, the reference wavelength is 630nm
7. All samples, washing liquids and various wastes should be treated as infectious agents. The stop solution is 2M sulfuric acid, and you must pay attention to safety when using it.
Storage conditions and validity period
1. Kit storage :; 2-8 ℃.
2. Validity: 6 months

Anomatic Advantages:

· High purity solvent free gases

· Does not change the appearance of the canister. 

· Fully replicates the profile of canisters

· Hydrophobic low surface energy system

· Fully licensed sustainable process 

· Environmentally friendly with a low carbon footprint.  

Aluminum anodization is an electrochemical process that enhances the strength, durability, and appearance of aluminum.  When aluminum is oxidized, super-strong oxide (anodic) layers [grow" on the surface of the metal. These new layers are highly resistant to wear and corrosion.

The exacting part to part consistency achieved with our unique anodizing system cannot be obtained with conventional hoist line systems.  Conventional anodizing systems present parts with varied coating thicknesses and voids formed due to air pocketing. 

A consistent uniform 6 µm anodic film thicknesses will protect  certain drug formulations and the canister from degradation.   Heavier film thicknesses of up to 12µm can be provided to meet your more demanding application.

Drug Delivery Components

Drug Delivery Components,Contact Down Force Components,Precision Device Measures Components,Mdi Direct Placement Transducer Components

Anomatic(suzhou)metal packaging co,ltd , https://www.anomaticpharma.com